Section 7: TSP command reference
Series 2600B System SourceMeter® Instrument Reference Manual
7-388
2600BS-901-01 Rev. C / August 2016
tsplink.trigger[N].clear()
This function clears the event detector for a trigger. This function is not available on the Models
2604B/2614B/2634B.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
tsplink.trigger[
N
].clear()
N
The trigger line (1 to 3)
Details
The trigger event detector enters the detected state when an event is detected.
tsplink.trigger[
N
].clear()
clears a trigger event detector, discards the history of the trigger
line, and clears the
tsplink.trigger[
N
].overrun
attribute.
Example
tsplink.trigger[2].clear()
Clears trigger event on synchronization line 2.
Also see
(on page 7-389)
(on page 7-391)
(on page 7-392)
(on page 7-394)
(on page 7-396)
tsplink.trigger[N].EVENT_ID
This constant identifies the number that is used for the trigger events. This constant is not available on the Models
2604B/2614B/2634B.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Constant
Yes
Usage
eventID
= tsplink.trigger[
N
].EVENT_ID
eventID
The trigger event number