![Keithley SourceMeter 2600B Series Скачать руководство пользователя страница 763](http://html2.mh-extra.com/html/keithley/sourcemeter-2600b-series/sourcemeter-2600b-series_reference-manual_4085220763.webp)
Series 2600B System SourceMeter® Instrument Reference Manual
Section 9: Frequently asked questions (FAQs)
2600BS-901-01 Rev. C / August 2016
9-3
How do I upgrade the firmware?
Do not turn off power or remove the USB flash drive until the upgrade process is complete.
From the front panel:
1. Copy the firmware upgrade file to a USB flash drive.
2. Verify that the upgrade file is in the root subdirectory of the flash drive and that it is the only
firmware file in that location.
3. Disconnect any input and output terminals that are attached to the instrument.
4. Turn on instrument power.
5. Insert the flash drive into the USB port on the front panel of the instrument.
6. From the instrument front panel, press the
MENU
key.
7. Under System, select
Manage
.
8. To upgrade to a newer version of firmware, select
Upgrade to New
.
9. To return to a previous version of firmware, select
Downgrade to Older
.
10. If the instrument is controlled remotely, a message is displayed. Select
Yes
to continue.
11. When the upgrade is complete, reboot the instrument.
A message is displayed while the upgrade is in progress.
For additional information about upgrading the firmware, see
(on page A-4).
How do I use the digital I/O port?
You can use the Series 2600B digital input/output with the trigger model or to control an external
digital circuit, such as a device handler used to perform binning operations. To control or configure
any of the six digital input/output lines, send commands to the Series 2600B over a remote interface.
To use the Series 2600B digital I/O in a trigger link system (TLINK), connect it using a Series
2600B-TLINK Trigger Link Cable and configure the Series 2600B digital input and output lines.
For more information about the Series 2600B digital I/O port, see
(on page 3-82).
How do I trigger other instruments?
You can use the Series 2600B digital input/output to control an external digital circuit, such as a
device handler used to perform binning operations. For more information about the Series 2600B
digital I/O port, see
(on page 3-82).
You can also use the digital I/O in a trigger link system (TLINK) using a Series 2600B-TLINK Trigger
Link Cable.
Another option is Keithley Instruments TSP-Link
®
, a high-speed trigger synchronization and
communication bus that you can use to connect multiple instruments in a master and subordinate
configuration. See TSP-Link System Expansion Interface for additional information.