Section 7: TSP command reference
Series 2600B System SourceMeter® Instrument Reference Manual
7-186
2600BS-901-01 Rev. C / August 2016
scriptVar.save()
This function saves the script to nonvolatile memory or to a USB flash drive.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
No
Usage
scriptVar
.save()
scriptVar
.save(
filename
)
scriptVar
The name of variable that references the script
filename
The file name to use when saving the script to a USB flash drive
Details
The
scriptVar
.save()
function saves a script to nonvolatile memory or a USB flash drive. The
root folder of the USB flash drive has the absolute path
/usb1/
.
If no
filename
is specified (the filename parameter is an empty string), the script is saved to internal
nonvolatile memory. Only a script with
filename
defined can be saved to internal nonvolatile
memory. If a
filename
is given, the script is saved to the USB flash drive.
If no
filename
is specified (the filename parameter is an empty string), the script is saved to internal
nonvolatile memory. Only a script with
filename
defined can be saved to internal nonvolatile
memory. If a
filename
is given, the script is saved to the USB flash drive.
You can add the file extension, but it is not required. The only allowed extension is
.tsp
(see
Example 2).
Example 1
test8.save()
Saves the script referenced by the variable
test8
to nonvolatile memory.
Example 2
test8.save("/usb1/myScript.tsp")
Saves the script referenced by the variable
test8
to a file named
myScript.tsp
on your
USB flash drive.
Also see
(on page 6-8)