Series 2600B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2600BS-901-01 Rev. C / August 2016
7-357
Bit
B7
B6
B5
B4
B3
B2
B1
B0
Binary value
0/1
0/1
0/1
0/1
0/1
0/1
0/1
0/1
Decimal
128
64
32
16
8
4
2
1
Weights
(2
7
)
(2
6
)
(2
5
)
(2
4
)
(2
3
)
(2
2
)
(2
1
)
(2
0
)
Bit
B15
B14
B13
B12
B11
B10
B9
B8
Binary value
0/1
0/1
0/1
0/1
0/1
0/1
0/1
0/1
Decimal
32,768
16,384
8,192
4,096
2,048
1,024
512
256
Weights
(2
15
)
(2
14
)
(2
13
)
(2
12
)
(2
11
)
(2
10
)
(2
9
)
(2
8
)
Example 1
enableRegister = status.system5.
status.system5.NODE60
status.system2.enable = enableRegister
Uses constants to set bits B1 and B4 of
the system summary 5 enable register.
Example 2
-- decimal 18 = binary 0000 0000 0001 0010
enableRegister = 18
status.system5.enable = enableRegister
Uses a decimal value to set bits B1 and
B4 of the system summary 5 enable
register.
Also see
(on page 7-353)
System summary and standard event registers
(on page E-7)
SweepILinMeasureV()
This KISweep factory script function performs a linear current sweep with voltage measured at every step (point).
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
No
Usage
SweepILinMeasureV(
smu, starti, stopi, stime, points
)
smu
System SourceMeter
®
instrument channel (for example,
smua
refers to SMU
channel A)
starti
Sweep start current in amperes
stopi
Sweep stop current in amperes
stime
Settling time in seconds; occurs after stepping the source and before performing a
measurement
points
Number of sweep points (must be
≥
2)
Details
Data for voltage measurements, current source values, and timestamps are stored in
smu
X
.nvbuffer1.
If all parameters are omitted when this function is called, this function is executed with the parameters
set to the default values.