Series 2600B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2600BS-901-01 Rev. C / August 2016
7-245
Details
Reading this attribute returns the output state of the source. Setting this attribute switches the output
of the source on or off.
When the output is switched on, the SMU sources either voltage or current, as set by
smu
X
.source.func
.
Setting this attribute to
smu
X
.OUTPUT_HIGH_Z
causes the output to turn off and go to the High Z
mode. If the
smu
X
.source.output
is read after setting this attribute to
smu
X
.OUTPUT_HIGH_Z
, it
returns
0
.
Example
smua.source.output = smua.OUTPUT_ON
Turns on SMU channel A source
output.
Also see
(on page 2-49)
(on page 7-237)
smuX.source.outputenableaction
This attribute controls output enable action of the source.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
SMU reset
Instrument reset
Recall setup
Saved setup
0 (smuX.OE_NONE)
Usage
outputAction
= smu
X
.source.outputenableaction
smu
X
.source.outputenableaction =
outputAction
outputAction
The output enable action of the source; set to one of the following values:
0
or
smu
X
.OE_NONE
: No action
1
or
smu
X
.OE_OUTPUT_OFF
: Turns the source output off
X
Source-measure unit (SMU) channel (for example,
smua.source.outputenableaction
applies to SMU channel A)