Series 2600B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2600BS-901-01 Rev. C / August 2016
7-247
smuX.source.rangeY
This attribute contains the source range.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW) Yes
SMU reset
Instrument reset
Recall setup
Saved setup
Voltage:
Models 2601B/2602B/2604B: 100e-3 (100 mV)
Models 2611B/2612B/2614B/2634B/2635B/2636B:
200e-3 (200 mV)
Current:
Models 2601B/2602B/2604B/2611B/2612B/2614B:
100e-9 (100 nA)
Models 2634B/2635B/2636B: 1e-9 (1 nA)
Usage
rangeValue
= smu
X
.source.range
Y
smu
X
.source.range
Y
=
rangeValue
rangeValue
Set to the maximum expected voltage or current to be sourced
X
Source-measure unit (SMU) channel (for example,
smua.measure.rangev
applies to SMU channel A)
Y
SMU source function (
v
= voltage,
i
= current)
Details
This attribute contains a value that sets the source-measure unit (SMU) to a fixed range large enough
to source the value. When read, the attribute contains the range the instrument is presently on when
in autorange.
Assigning a value to this attribute sets the SMU to a fixed range large enough to source the assigned
value. The instrument selects the best range for sourcing a value of
rangeValue
.
Reading this attribute returns the positive full-scale value of the source range the SMU is currently
using. With source autoranging enabled, the output level controls the range. Querying the range after
the level is set returns the range the instrument chose as appropriate for that source level.
This attribute is primarily intended to eliminate the time required by the automatic range selection
performed by a sourcing instrument. Because selecting a fixed range prevents autoranging, an
overrange condition can occur. For example, sourcing 10.0 V on the Model 2601B/2602B/2604B 6 V
range or sourcing 5.0 V on the Model 2611B/2612B/2614B 2 V range causes an overrange condition.
Example
smua.source.rangev = 1
Selects the 1 V source range for SMU channel A.