Series 2600B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2600BS-901-01 Rev. C / August 2016
7-217
Example
smua.measure.calibratev(1, 1e-4, 1e-5, 0.92,
0.903)
SMU channel A calibrates voltage
measurement using following values: 1 V
calibration range, 1e
−
4 for +zero
measurement reading, 1e
−
5 for +zero
DMM measurement reading, 0.92 for
+
FS measurement reading, and 0.903 for
the
+
FS DMM measurement reading.
Also see
(on page B-18)
(on page 7-201)
(on page 7-203)
(on page 7-204)
(on page 7-235)
smuX.measure.count
This attribute sets the number of measurements performed when a measurement is requested.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
SMU reset
Recall setup
Saved setup
1
Usage
count
= smu
X
.measure.count
smu
X
.measure.count =
count
count
Number of measurements
X
Source-measure unit (SMU) channel (for example,
smua.measure.count
applies
to SMU channel A)
Details
This attribute controls the number of measurements taken any time a measurement is requested.
When using a reading buffer with a measure command, this attribute also controls the number of
readings to be stored.
If
smu
X
.measure.count
is set to a value greater than 1, any measurement delay set by
smu
X
.measure.delay
will only occur before the first measurement, while the
smu
X
.measure.interval
controls the interval between successive measurements.
Example
smua.measure.count = 10
Sets the SMU channel A measure count to 10.
Also see
(on page 7-173)
(on page 7-194)
(on page 7-218)
(on page 7-222)
(on page 7-232)