Section 7: TSP command reference
Series 2600B System SourceMeter® Instrument Reference Manual
7-378
2600BS-901-01 Rev. C / August 2016
Details
When pass-through mode is enabled, triggers are passed through immediately and initiate the delay.
When disabled, a trigger only initiates a delay.
Example
trigger.timer[1].passthrough = true
Enables pass-through mode on trigger timer 1.
Also see
(on page 7-378)
trigger.timer[N].reset()
This function resets some of the trigger timer settings to their factory defaults.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
trigger.timer[
N
].reset()
N
Trigger timer number (1 to 8)
Details
The
trigger.timer[
N
].reset()
function resets the following attributes to their factory defaults:
•
trigger.timer[
N
].count
•
trigger.timer[
N
].delay
•
trigger.timer[
N
].delaylist
•
trigger.timer[
N
].passthrough
•
trigger.timer[
N
].stimulus
It also clears
trigger.timer[
N
].overrun
.
Example
trigger.timer[1].reset()
Resets the attributes associated with timer 1 back
to factory default values.
Also see
(on page 7-374)
(on page 7-375)
(on page 7-377)
(on page 7-379)