Series 2600B System SourceMeter® Instrument Reference Manual
Appendix B: Calibration
2600BS-901-01 Rev. C / August 2016
B-39
Step 8. Lock out calibration
To lock out further calibration, send the following command after completing the calibration
procedure:
smua.cal.lock()
Step 8. Repeat calibration procedure for Model
2602B/2604B/2612B/2614B/2634B/2636B Channel B
For the Models 2602B/2604B/2612B/2614B/2634B/2636B only, repeat the entire procedure above for
Channel B. Be sure to:
•
Make test connections to Channel B terminals.
•
Substitute "
smub
" for "
smua
" in all commands.