Series 2600B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2600BS-901-01 Rev. C / August 2016
7-317
A set bit indicates the specified timer generated an action overrun because it was still processing a
delay from a previous trigger when a new trigger was received.
For information about .condition, .enable, .event, .ntr, and .ptr registers, refer to
Enable and transition registers
(on page E-19). The individual bits of this
register are defined in the following table.
Bit
Value
Description
B0
Not used
Not applicable
B1
status.operation.instrument.trigger_timer.trigger_overrun.TMR1
Bit B1 decimal
value: 2
B2
status.operation.instrument.trigger_timer.trigger_overrun.TMR2
Bit B2 decimal
value: 4
B3
status.operation.instrument.trigger_timer.trigger_overrun.TMR3
Bit B3 decimal
value: 8
B4
status.operation.instrument.trigger_timer.trigger_overrun.TMR4
Bit B4 decimal
value: 16
B5
status.operation.instrument.trigger_timer.trigger_overrun.TMR5
Bit B5 decimal
value: 32
B6
status.operation.instrument.trigger_timer.trigger_overrun.TMR6
Bit B6 decimal
value: 64
B7
status.operation.instrument.trigger_timer.trigger_overrun.TMR7
Bit B7 decimal
value: 128
B8
status.operation.instrument.trigger_timer.trigger_overrun.TMR8
Bit B8 decimal
value: 256
B9-B15 Not used
Not applicable
As an example, to set bit B1 of the operation status trigger timer trigger overrun enable register, set
status.operation.instrument.trigger_timer.trigger_overrun.enable =
status.operation.instrument.trigger_timer.trigger_overrun.TMR1
.
In addition to the above constants,
operationRegister
can be set to the numeric equivalent of the
bit to set. To set more than one bit of the register, set
operationRegister
to the sum of their
decimal weights. For example, to set bits B1 and B4, set
operationRegister
to 18 (which is the
sum of 2 + 16).
Bit
B7
B6
B5
B4
B3
B2
B1
B0
Binary value
0/1
0/1
0/1
0/1
0/1
0/1
0/1
0/1
Decimal
128
64
32
16
8
4
2
1
Weights
(2
7
)
(2
6
)
(2
5
)
(2
4
)
(2
3
)
(2
2
)
(2
1
)
(2
0
)