Series 2600B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2600BS-901-01 Rev. C / August 2016
7-269
Also see
(on page 7-266)
(on page 7-269)
(on page 7-270)
smuX.trigger.source.listY()
This function configures an array-based source sweep.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
smu
X
.trigger.source.list
Y
(
sweepList
)
X
Source-measure unit (SMU) channel (for example,
smua.trigger.source.listv({5})
applies to SMU channel A)
Y
SMU source function (
v
= voltage,
i
= current)
sweepList
An array of source values
Details
This function configures the source action to be a list sweep in a subsequent sweep. During the
sweep, the source will output the sequence of source values given in the
sweepList
array.
If the subsequent sweep has more points than specified in
sweepList
, the source will restart at the
beginning. This means that if the trigger count is greater than the number of points in a sweep as
configured, the SMU will satisfy the trigger count by restarting the sweep values from the beginning.
If the subsequent sweep has fewer points than specified in
sweepList
, the extra values will be
ignored. This means that if the trigger count is less than the number of source values configured, the
SMU will satisfy the trigger count and ignore the remaining source values.
In cases where the first sweep point is a nonzero value, it may be necessary to pre-charge the circuit
so that the sweep will return a stable value for the first measured point without penalizing remaining
points in the sweep.
The SMU will only store the most recent configured source action. The last call to
smu
X
.trigger.source.linear
Y
()
,
smu
X
.trigger.source.list
Y
()
, or
smu
X
.trigger.source.log
Y
()
is used for the source action.
Source functions cannot be changed within a sweep.
After configuring the sweep source values, enable the source action by setting
smu
X
.trigger.source.action
.
Example
smua.trigger.source.listv({3, 1, 4, 5, 2})
Sweeps SMU channel A through
3 V, 1 V, 4 V, 5 V, and 2 V.