smuX.cal.due
This attribute stores the calibration due date for the next calibration.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
SMU cal. restore
SMU nonvolatile
memory
0
Usage
calDue
= smu
X
.cal.due
smu
X
.cal.due =
calDue
calDue
Due date of next calibration (
0
indicates that no date is set)
X
Source-measure unit (SMU) channel (for example,
smua.cal.due
applies to
SMU channel A)
Details
This attribute stores the calibration due date associated with the active calibration set. The calibration
due date can be read at any time but can only be assigned a new value when calibration has been
enabled with the
smu
X
.cal.unlock()
function. It is typically set to the date when the next
calibration should be performed.
This attribute is stored with the active calibration set. If a different calibration set is restored, this
attribute will reflect the due date stored with that set.
smu
X
.cal.due
must be set to the date the next calibration is required using the UTC time and date.
The date is stored as the number of seconds since UTC 12:00 am Jan 1, 1970.
Due to the internal storage format,
smu
X
.cal.due
is only accurate to within a few minutes of the
value set.
Example
smua.cal.due = os.time() + 365 * 24 * 60 * 60
Sets the SMU channel A calibration due date
equal to one year from the current time set on
the instrument.