Section 7: TSP command reference
Series 2600B System SourceMeter® Instrument Reference Manual
7-312
2600BS-901-01 Rev. C / August 2016
Usage
operationRegister
=
status.operation.instrument.trigger_blender.trigger_overrun.condition
operationRegister
=
status.operation.instrument.trigger_blender.trigger_overrun.enable
operationRegister
=
status.operation.instrument.trigger_blender.trigger_overrun.event
operationRegister
=
status.operation.instrument.trigger_blender.trigger_overrun.ntr
operationRegister
=
status.operation.instrument.trigger_blender.trigger_overrun.ptr
status.operation.instrument.trigger_blender.trigger_overrun.enable =
operationRegister
status.operation.instrument.trigger_blender.trigger_overrun.ntr =
operationRegister
status.operation.instrument.trigger_blender.trigger_overrun.ptr =
operationRegister
operationRegister
The status of the operation status trigger blender overrun register; a zero (0)
indicates no bits set (also send 0 to clear all bits); other values indicate various bit
settings
Details
These attributes are used to read or write to the operation status trigger blender overrun registers.
Reading a status register returns a value. The binary equivalent of the returned value indicates which
register bits are set. The least significant bit of the binary number is bit B0, and the most significant bit
is bit B15. For example, if a value of
18
is read as the value of the condition register, the binary
equivalent is 0000 0000 0001 0010. This value indicates that bit B1 and bit B4 are set.
B15 B14 B13 B12 B11 B10 B9
B8
B7
B6
B5
B4
B3
B2
B1
B0
**
>
>
>
>
>
>
>
>
>
>
>
>
>
>
*
0
0
0
0
0
0
0
0
0
0
0
1
0
0
1
0
* Least significant bit