Phaco Power Settings
Z370584 Rev. D
6-20
Program Settings
Continuous Phaco Power
Continuous phaco power delivers continuous, uninterrupted phaco power to the handpiece
and requires no pulse rate setting.
Pulse Mode
Pulse Mode delivers phaco in pulses of 1 to 100 ms when the foot pedal is in position 3. You
can set this in a range of 1 to 100 pulses per second (pps). The default setting is 20 pps.
To set the Pulse Mode range:
1.
Tap the
Pulse Mode
button.
2. Tap the
pulse setting
number. A Settings dialog box opens.
3. Tap the up or down arrows to increase or decrease the pps from 1 to 100.
4. Tap
Finished
to close the screen.
Short Pulse Phaco Power
Short pulse delivers phaco in pulses of 50 ms when the foot pedal is in position 3. You can set
this in a range of 1 to 14 pulses per second (pps). The actual number of pps is on the button to
the right of the Short Pulse button.
To set the Short Pulse range:
1.
Tap the
Short Pulse
button.
2. Tap on the
pulse setting
number. A Settings dialog box opens.
3. Tap the up or down arrows to increase or decrease the pps from 1 to 14.
4. Tap
Finished
to close the screen.
Long Pulse Phaco Power
Long Pulse delivers phaco in pulses of 150 ms when the foot pedal is in position 3. You can
set this in a range of 1 to 6 pulses per second (pps). The actual number of pps is on the button
to the right of the Long Pulse button.
To set the Long Pulse range:
1.
Tap the
Long Pulse
button.
2. Tap the
pulse setting
number. A Settings dialog box opens.
3. Tap the up or down arrows to increase or decrease the pps from 1 to 6.
4. Tap
Finished
to close the screen.
Low Power Pulse Phaco Power
Low Power Pulse generates short pulses of ultrasonic power in foot pedal position 3. When
you press the foot pedal, the pulses become longer and eventually blend together to become
continuous phaco power.
Содержание Veritas
Страница 1: ...Z370584 Rev D 09 2020 Johnson Johnson Surgical Vision Inc 2020 Software Release 8 000 OPERATOR MANUAL...
Страница 2: ...Z370584 Rev D This page was intentionally left blank...
Страница 4: ...Z370584 Rev D This page was intentionally left blank...
Страница 22: ...Z370584 Rev D 1 12 This page was intentionally left blank...
Страница 36: ...Z370584 Rev D 2 14 This page was intentionally left blank...
Страница 54: ...Z370584 Rev D 3 18 This page was intentionally left blank...
Страница 122: ...Z370584 Rev D 5 42 This page was intentionally left blank...
Страница 124: ...Create a New Program Z370584 Rev D 6 2 Program Settings Figure 6 1 Surgeons and Programs Screen 1 On Select Program tap...
Страница 146: ...WHITESTAR Technology Z370584 Rev D 6 24 Program Settings Figure 6 14 Duty Cycles for Variable WS...
Страница 158: ...Z370584 Rev D 6 36 This page was intentionally left blank...
Страница 184: ...Z370584 Rev D 7 26 This page was intentionally left blank...
Страница 198: ...Z370584 Rev D 9 4 This page was intentionally left blank...
Страница 270: ...Z370584 Rev D I 8 This page was intentionally left blank...