Error Messages, Troubleshooting, and Diagnostics
Z370584 Rev. D
Status, Warning, and Error Messages
10-19
3. If the error does not clear, select
End Case
.
4. Select
Shutdown
.
5. Select
Yes
and wait for the system to power off.
6. Start up the system.
7. Retune the handpiece.
8. If the error still does not clear, replace the handpiece with a new handpiece and tune the
handpiece.
9. If the error still does not clear or if the error continues to occur, document the error
message and contact JJSV for technical service.
206 Phaco controller ADC error.
Probable cause: Phaco controller ADC problem.
Corrective action:
1.
Select
End Case
.
2. Select
Shutdown
.
3. Select
Yes
and wait for the system to power off.
4. Start up the system.
5. If the error still does not clear or if the error continues to occur, document the error
message and contact JJSV for technical service.
207 Handpiece not supported.
Probable cause:
•
Incompatible handpiece.
Corrective action:
1.
Change to a different handpiece.
2. Tune the handpiece and continue surgery.
3. If the error still does not clear or if the error continues to occur, document the error
message and contact JJSV for technical service.
208 Handpiece error.
Probable cause:
•
Hardware failure.
•
The handpiece is shorting out.
Corrective action:
1.
Check the handpiece connections
2. Retune the handpiece.
3. If the error does not clear, select
End Case
.
Содержание Veritas
Страница 1: ...Z370584 Rev D 09 2020 Johnson Johnson Surgical Vision Inc 2020 Software Release 8 000 OPERATOR MANUAL...
Страница 2: ...Z370584 Rev D This page was intentionally left blank...
Страница 4: ...Z370584 Rev D This page was intentionally left blank...
Страница 22: ...Z370584 Rev D 1 12 This page was intentionally left blank...
Страница 36: ...Z370584 Rev D 2 14 This page was intentionally left blank...
Страница 54: ...Z370584 Rev D 3 18 This page was intentionally left blank...
Страница 122: ...Z370584 Rev D 5 42 This page was intentionally left blank...
Страница 124: ...Create a New Program Z370584 Rev D 6 2 Program Settings Figure 6 1 Surgeons and Programs Screen 1 On Select Program tap...
Страница 146: ...WHITESTAR Technology Z370584 Rev D 6 24 Program Settings Figure 6 14 Duty Cycles for Variable WS...
Страница 158: ...Z370584 Rev D 6 36 This page was intentionally left blank...
Страница 184: ...Z370584 Rev D 7 26 This page was intentionally left blank...
Страница 198: ...Z370584 Rev D 9 4 This page was intentionally left blank...
Страница 270: ...Z370584 Rev D I 8 This page was intentionally left blank...