Status, Warning, and Error Messages
Z370584 Rev. D
10-28
Error Messages, Troubleshooting, and Diagnostics
9. Replace the foot pedal.
10. If the error still does not clear or if the error continues to occur, document the error
message and contact JJSV for technical service.
420 Foot pedal battery too low.
Probable cause: Foot pedal battery to low to operate wireless.
Corrective action:
1.
Connect the foot pedal to the system with the foot pedal cable to clear the error.
2. Continue with the current case once the error is cleared.
501 Prime excessive vacuum error.
Probable cause:
•
The system does not prime.
•
There is too much vacuum.
Corrective action:
1.
Prime the system.
2. Reinstall the Fluidics Tubing pack and prime.
3. Check the handpiece. Replace the tip if needed.
4. If the error still does not clear, replace the handpiece with a new handpiece, and tune the
handpiece.
5. If the error does not clear, replace the Fluidics Tubing pack with a new Fluidics Tubing
pack and prime the system.
6. If the error does not clear or if the error continues to occur, document the error message
and contact JJSV for technical service.
502 Prime low bottle height error.
Probable cause: The bottle is not at the proper height.
Corrective action:
1.
Increase the bottle height.
2. Prime the system.
3. Tap the drip chamber to ensure the valve operates properly.
4. If the error still does not clear, replace the Fluidics Tubing pack with a new Fluidics
Tubing pack and prime the system.
5. If the error does not clear or if the error continues to occur, document the error message
and contact JJSV for technical service.
503 Prime low vacuum error.
Probable cause: Hardware failure.
Содержание Veritas
Страница 1: ...Z370584 Rev D 09 2020 Johnson Johnson Surgical Vision Inc 2020 Software Release 8 000 OPERATOR MANUAL...
Страница 2: ...Z370584 Rev D This page was intentionally left blank...
Страница 4: ...Z370584 Rev D This page was intentionally left blank...
Страница 22: ...Z370584 Rev D 1 12 This page was intentionally left blank...
Страница 36: ...Z370584 Rev D 2 14 This page was intentionally left blank...
Страница 54: ...Z370584 Rev D 3 18 This page was intentionally left blank...
Страница 122: ...Z370584 Rev D 5 42 This page was intentionally left blank...
Страница 124: ...Create a New Program Z370584 Rev D 6 2 Program Settings Figure 6 1 Surgeons and Programs Screen 1 On Select Program tap...
Страница 146: ...WHITESTAR Technology Z370584 Rev D 6 24 Program Settings Figure 6 14 Duty Cycles for Variable WS...
Страница 158: ...Z370584 Rev D 6 36 This page was intentionally left blank...
Страница 184: ...Z370584 Rev D 7 26 This page was intentionally left blank...
Страница 198: ...Z370584 Rev D 9 4 This page was intentionally left blank...
Страница 270: ...Z370584 Rev D I 8 This page was intentionally left blank...