Program Settings
Z370584 Rev. D
6-8
Program Settings
Phaco
1.
Tap
to open the Settings screen.
2. Select a Phaco submode on the left (
).
Figure 6-6 — Phaco Settings Screen
3. If desired, tap the submode name at the top of the screen (outlined in red in
customize the submode name.
4. Select the pump type (
Peristaltic
or
Venturi
).
5. Use the up and down arrows to set the
Aspiration
rate (10 to 60).
6. Use the up and down arrows to set the
Vacuum
. If you are using Venturi, the range is 0 to
600. If you are using Peristaltic, the range is 0 to 650.
7. Use the up and down arrows to set the
Power
(0% to 100%).
8. Select the delivery mode (
Linear
or
Panel
).
9. Set the
Bottle Height
(Advanced Fluidics) or
Total Height
(Advanced Infusion).
10. Tap
Save
to save the program.
For additional information see
“FUSION Fluidics Phaco” on page 6-27
and
Содержание Veritas
Страница 1: ...Z370584 Rev D 09 2020 Johnson Johnson Surgical Vision Inc 2020 Software Release 8 000 OPERATOR MANUAL...
Страница 2: ...Z370584 Rev D This page was intentionally left blank...
Страница 4: ...Z370584 Rev D This page was intentionally left blank...
Страница 22: ...Z370584 Rev D 1 12 This page was intentionally left blank...
Страница 36: ...Z370584 Rev D 2 14 This page was intentionally left blank...
Страница 54: ...Z370584 Rev D 3 18 This page was intentionally left blank...
Страница 122: ...Z370584 Rev D 5 42 This page was intentionally left blank...
Страница 124: ...Create a New Program Z370584 Rev D 6 2 Program Settings Figure 6 1 Surgeons and Programs Screen 1 On Select Program tap...
Страница 146: ...WHITESTAR Technology Z370584 Rev D 6 24 Program Settings Figure 6 14 Duty Cycles for Variable WS...
Страница 158: ...Z370584 Rev D 6 36 This page was intentionally left blank...
Страница 184: ...Z370584 Rev D 7 26 This page was intentionally left blank...
Страница 198: ...Z370584 Rev D 9 4 This page was intentionally left blank...
Страница 270: ...Z370584 Rev D I 8 This page was intentionally left blank...