System Configuration
Z370584 Rev. D
Service Interval
7-17
Service Interval
The Service Interval screen provides information on the previous and upcoming date for
Preventative Maintenance, as well as system usage count information.
1.
On the main screen, tap
Configuration
.
2. Tap
Service Interval
.
Figure 7-15 — Service Interval Screen
3. Tap
Finished
to close the dialog box.
4. Tap
Exit
to close the Configuration screen.
Import/Export Database
Export
Export can be used to export surgeon program(s) to a USB drive.
1.
From Configuration, tap
Import/Export Database
.
NOTE: Ensure you place the USB device into the USB port on the back of the system.
Содержание Veritas
Страница 1: ...Z370584 Rev D 09 2020 Johnson Johnson Surgical Vision Inc 2020 Software Release 8 000 OPERATOR MANUAL...
Страница 2: ...Z370584 Rev D This page was intentionally left blank...
Страница 4: ...Z370584 Rev D This page was intentionally left blank...
Страница 22: ...Z370584 Rev D 1 12 This page was intentionally left blank...
Страница 36: ...Z370584 Rev D 2 14 This page was intentionally left blank...
Страница 54: ...Z370584 Rev D 3 18 This page was intentionally left blank...
Страница 122: ...Z370584 Rev D 5 42 This page was intentionally left blank...
Страница 124: ...Create a New Program Z370584 Rev D 6 2 Program Settings Figure 6 1 Surgeons and Programs Screen 1 On Select Program tap...
Страница 146: ...WHITESTAR Technology Z370584 Rev D 6 24 Program Settings Figure 6 14 Duty Cycles for Variable WS...
Страница 158: ...Z370584 Rev D 6 36 This page was intentionally left blank...
Страница 184: ...Z370584 Rev D 7 26 This page was intentionally left blank...
Страница 198: ...Z370584 Rev D 9 4 This page was intentionally left blank...
Страница 270: ...Z370584 Rev D I 8 This page was intentionally left blank...