Status, Warning, and Error Messages
Z370584 Rev. D
10-20
Error Messages, Troubleshooting, and Diagnostics
4. Select
Shutdown
.
5. Select
Yes
and wait for the system to power off.
6. Start up the system.
7. Retune the handpiece.
8. If the error still does not clear, replace the handpiece with a new handpiece, and tune the
handpiece.
9. If the error still does not clear or if the error continues to occur, document the error
message and contact JJSV for technical service.
210 Phaco RAM error.
Probable cause: Bad Microcontroller.
Corrective action:
1.
Select
End Case
.
2. Select
Shutdown
.
3. Select
Yes
and wait for the system to power off.
4. Start up the system.
5. If the error does not clear or if the error continues to occur, document the error message
and contact JJSV for technical service.
211 Phaco ROM error.
Probable cause: Bad Microcontroller.
Corrective action:
1.
Select
End Case
.
2. Select
Shutdown
.
3. Select
Yes
and wait for the system to power off.
4. Start up the system.
5. If the error does not clear or if the error continues to occur, document the error message
and contact JJSV for technical service.
212 Phaco controller timeout.
Probable cause: Instrument Host failure or bad Microcontroller.
Corrective action:
1.
Select
End Case
.
2. Select
Shutdown
.
3. Select
Yes
and wait for the system to power off.
4. Start up the system.
5. If the error does not clear or if the error continues to occur, document the error message
and contact JJSV for technical service.
Содержание Veritas
Страница 1: ...Z370584 Rev D 09 2020 Johnson Johnson Surgical Vision Inc 2020 Software Release 8 000 OPERATOR MANUAL...
Страница 2: ...Z370584 Rev D This page was intentionally left blank...
Страница 4: ...Z370584 Rev D This page was intentionally left blank...
Страница 22: ...Z370584 Rev D 1 12 This page was intentionally left blank...
Страница 36: ...Z370584 Rev D 2 14 This page was intentionally left blank...
Страница 54: ...Z370584 Rev D 3 18 This page was intentionally left blank...
Страница 122: ...Z370584 Rev D 5 42 This page was intentionally left blank...
Страница 124: ...Create a New Program Z370584 Rev D 6 2 Program Settings Figure 6 1 Surgeons and Programs Screen 1 On Select Program tap...
Страница 146: ...WHITESTAR Technology Z370584 Rev D 6 24 Program Settings Figure 6 14 Duty Cycles for Variable WS...
Страница 158: ...Z370584 Rev D 6 36 This page was intentionally left blank...
Страница 184: ...Z370584 Rev D 7 26 This page was intentionally left blank...
Страница 198: ...Z370584 Rev D 9 4 This page was intentionally left blank...
Страница 270: ...Z370584 Rev D I 8 This page was intentionally left blank...