WHITESTAR Technology
Z370584 Rev. D
6-22
Program Settings
Table 6.7 – WHITESTAR Technology Parameter Settings
Figure 6-12 — Duty Cycle
Setting
Pulse On Time (ms)
Pulse Off Time (ms)
Duty Cycle
Pulse Rate (pps)
6/12
6
12
33%
55
4/8
4
8
33%
83
6/4
6
4
60%
100
6/8
6
8
43%
71
8/4
8
4
67%
83
4/24
4
24
14%
35
6/28
6
28
18%
29
6/24
6
24
20%
33
8/24
8
24
25%
31
6/18
6
18
25%
41
Содержание Veritas
Страница 1: ...Z370584 Rev D 09 2020 Johnson Johnson Surgical Vision Inc 2020 Software Release 8 000 OPERATOR MANUAL...
Страница 2: ...Z370584 Rev D This page was intentionally left blank...
Страница 4: ...Z370584 Rev D This page was intentionally left blank...
Страница 22: ...Z370584 Rev D 1 12 This page was intentionally left blank...
Страница 36: ...Z370584 Rev D 2 14 This page was intentionally left blank...
Страница 54: ...Z370584 Rev D 3 18 This page was intentionally left blank...
Страница 122: ...Z370584 Rev D 5 42 This page was intentionally left blank...
Страница 124: ...Create a New Program Z370584 Rev D 6 2 Program Settings Figure 6 1 Surgeons and Programs Screen 1 On Select Program tap...
Страница 146: ...WHITESTAR Technology Z370584 Rev D 6 24 Program Settings Figure 6 14 Duty Cycles for Variable WS...
Страница 158: ...Z370584 Rev D 6 36 This page was intentionally left blank...
Страница 184: ...Z370584 Rev D 7 26 This page was intentionally left blank...
Страница 198: ...Z370584 Rev D 9 4 This page was intentionally left blank...
Страница 270: ...Z370584 Rev D I 8 This page was intentionally left blank...