Wireless Foot Pedal Setup
Z370584 Rev. D
7-8
System Configuration
Figure 7-7 — Wireless Pairing Instructions
7. The system communicates the appropriate data to the foot pedal. The system displays a
progress screen during the pairing process.
8. When pairing completes, the progress screen closes and the system displays the
message: “The foot pedal is now paired with the system”.
If the blue light is on, the wireless foot pedal is paired.
If the blue light is blinking the wireless foot pedal is not paired.
If the green light is on, the batteries in the foot pedal are charged.
If the green light is blinking, the batteries in the wireless foot pedal need to be charged.
See
“Charging Options for Wireless Foot Pedals” on page 10-4
9. Test the foot pedal or tap
Finished
to close the test screen.
10. Tap
Exit
to close the Configuration screen.
NOTE: If you have problems pairing the Advanced Control Pedal (ACP), remove any other
Bluetooth devices from the area. Multiple Bluetooth devices can affect the pairing
process.
Wired Pairing
1.
On the main screen, tap
Configuration
.
2. Tap
Wireless Setup
.
3. On the Wireless Setup dialog box, tap the
Advanced Control Pedal
button (
).
4. On the Foot Pedal Pairing screen, tap
Wired Pairing
). The system displays the
message: “Ensure the cable is attached to the foot pedal”.
5. Tap
OK
.
6. The system communicates the appropriate data to the foot pedal. When the system
receives the appropriate data back from the foot pedal the system displays the message:
“Remove the cable from the foot pedal”. The system displays a progress screen during
the pairing process.
7. When pairing completes, the progress screen closes and the system displays the
message: “The foot pedal is now paired with the system”.
8. Test the foot pedal or tap
Finished
to close the test screen.
9. Tap
Exit
to close the Configuration screen.
Содержание Veritas
Страница 1: ...Z370584 Rev D 09 2020 Johnson Johnson Surgical Vision Inc 2020 Software Release 8 000 OPERATOR MANUAL...
Страница 2: ...Z370584 Rev D This page was intentionally left blank...
Страница 4: ...Z370584 Rev D This page was intentionally left blank...
Страница 22: ...Z370584 Rev D 1 12 This page was intentionally left blank...
Страница 36: ...Z370584 Rev D 2 14 This page was intentionally left blank...
Страница 54: ...Z370584 Rev D 3 18 This page was intentionally left blank...
Страница 122: ...Z370584 Rev D 5 42 This page was intentionally left blank...
Страница 124: ...Create a New Program Z370584 Rev D 6 2 Program Settings Figure 6 1 Surgeons and Programs Screen 1 On Select Program tap...
Страница 146: ...WHITESTAR Technology Z370584 Rev D 6 24 Program Settings Figure 6 14 Duty Cycles for Variable WS...
Страница 158: ...Z370584 Rev D 6 36 This page was intentionally left blank...
Страница 184: ...Z370584 Rev D 7 26 This page was intentionally left blank...
Страница 198: ...Z370584 Rev D 9 4 This page was intentionally left blank...
Страница 270: ...Z370584 Rev D I 8 This page was intentionally left blank...