Status, Warning, and Error Messages
Z370584 Rev. D
10-18
Error Messages, Troubleshooting, and Diagnostics
9. If the error still does not clear or if the error continues to occur, document the error
message and contact JJSV for technical service.
202 Phaco power error.
Probable cause: Hardware failure.
Corrective action:
1.
Check the handpiece connections.
2. Retune the handpiece.
3. If the error does not clear, select
End Case
.
4. Select
Shutdown
.
5. Select
Yes
and wait for the system to power off.
6. Start up the system.
7. Retune the handpiece.
8. If the error still does not clear, replace the handpiece with a new handpiece and tune the
handpiece.
9. If the error still does not clear or if the error continues to occur, document the error
message and contact JJSV for technical service.
203 Phaco handpiece error.
Probable cause: The handpiece is bad.
Corrective action:
1.
Check the handpiece connections.
2. Retune the handpiece.
3. If the error does not clear, select
End Case
.
4. Select
Shutdown
.
5. Select
Yes
and wait for the system to power off.
6. Start up the system.
7. Retune the handpiece.
8. If the error still does not clear, replace the handpiece with a new handpiece and tune the
handpiece.
9. If the error still does not clear or if the error continues to occur, document the error
message and contact JJSV for technical service.
204 Phaco handpiece error.
Probable cause: A wire in the handpiece is broken.
Corrective action:
1.
Check the handpiece connections.
2. Retune the handpiece.
Содержание Veritas
Страница 1: ...Z370584 Rev D 09 2020 Johnson Johnson Surgical Vision Inc 2020 Software Release 8 000 OPERATOR MANUAL...
Страница 2: ...Z370584 Rev D This page was intentionally left blank...
Страница 4: ...Z370584 Rev D This page was intentionally left blank...
Страница 22: ...Z370584 Rev D 1 12 This page was intentionally left blank...
Страница 36: ...Z370584 Rev D 2 14 This page was intentionally left blank...
Страница 54: ...Z370584 Rev D 3 18 This page was intentionally left blank...
Страница 122: ...Z370584 Rev D 5 42 This page was intentionally left blank...
Страница 124: ...Create a New Program Z370584 Rev D 6 2 Program Settings Figure 6 1 Surgeons and Programs Screen 1 On Select Program tap...
Страница 146: ...WHITESTAR Technology Z370584 Rev D 6 24 Program Settings Figure 6 14 Duty Cycles for Variable WS...
Страница 158: ...Z370584 Rev D 6 36 This page was intentionally left blank...
Страница 184: ...Z370584 Rev D 7 26 This page was intentionally left blank...
Страница 198: ...Z370584 Rev D 9 4 This page was intentionally left blank...
Страница 270: ...Z370584 Rev D I 8 This page was intentionally left blank...