Introduction
Z370584 Rev. D
Device Description
1-7
automatically turns off can be set per surgeon preference. Please note that when this feature
is enabled, a significantly large wound leakage or air bubbles in the tubing lines may affect
the continuous irrigation auto-off.
Advanced Infusion
Advanced Infusion is available when the VERITAS Advanced Infusion Pack is installed and
the feature is selected in the Program Settings. When this feature is enabled, pressurized air
(equivalent to 30 cm of bottle height) is added to the irrigation source. The pressurized air is
supplied by an additional tubing line that is connected to the side panel of the console during
pack setup.
NOTE: Using the Advanced Infusion feature in conjunction with vitrectomy at altitudes
above 5,000 feet may result in lower than expected irrigation pressure.
Programmable Operating Parameters
The system is programmable through the screens on the touchscreen monitor. You can
select your desired settings for each portion of the anterior surgical procedure. The
instrument program memory stores up to 100 surgeon names with a maximum of 20 different
setups, plus the two JJSV default settings programs—Advanced Fluidics Program and
Advanced Infusion Program. This allows different users to preset their preferences, or
individual users to select setups for different procedures, including a personalized initial
operating mode.
MMP – Multiple Mode Programming
Multiple submodes are available within the system operating modes. The multiple submodes
allows you to preset your settings for specific techniques such as phaco chop or viscoelastic
removal.
Programmable IV Pole
The system has a programmable IV pole. You can set the programmable IV pole height
parameters independently for each of the Phaco modes plus settings for I/A and vitrectomy.
During surgery, the programmable IV pole height changes to the preprogrammed height
when you switch modes.
The automated and programmable IV pole allows adjustment of the bottle height to provide
gravity infusion through each procedural phase. The up and down arrows on the touchscreen
or the wireless VERITAS Remote Control raises or lowers the irrigation source that contains
the bottled balanced salt solution, while maintaining the sterility within the operating field. A
separate up and down switch allows IV pole adjustment from the side of the system. The foot
pedal can also be programmed to move the IV pole.
NOTE: The IV pole movement is calibrated in centimeters. When the IV pole height is set to
inches, the height shown may not update immediately upon pressing the up arrow or
down arrow.
Содержание Veritas
Страница 1: ...Z370584 Rev D 09 2020 Johnson Johnson Surgical Vision Inc 2020 Software Release 8 000 OPERATOR MANUAL...
Страница 2: ...Z370584 Rev D This page was intentionally left blank...
Страница 4: ...Z370584 Rev D This page was intentionally left blank...
Страница 22: ...Z370584 Rev D 1 12 This page was intentionally left blank...
Страница 36: ...Z370584 Rev D 2 14 This page was intentionally left blank...
Страница 54: ...Z370584 Rev D 3 18 This page was intentionally left blank...
Страница 122: ...Z370584 Rev D 5 42 This page was intentionally left blank...
Страница 124: ...Create a New Program Z370584 Rev D 6 2 Program Settings Figure 6 1 Surgeons and Programs Screen 1 On Select Program tap...
Страница 146: ...WHITESTAR Technology Z370584 Rev D 6 24 Program Settings Figure 6 14 Duty Cycles for Variable WS...
Страница 158: ...Z370584 Rev D 6 36 This page was intentionally left blank...
Страница 184: ...Z370584 Rev D 7 26 This page was intentionally left blank...
Страница 198: ...Z370584 Rev D 9 4 This page was intentionally left blank...
Страница 270: ...Z370584 Rev D I 8 This page was intentionally left blank...