Program Settings
Z370584 Rev. D
Program Settings
6-9
Irrigation and Aspiration (I/A)
1.
Tap
to open the Settings screen.
2. Select an Irrigation/Aspiration submode on the left (
Figure 6-7 — Irrigation/Aspiration Settings
3. If desired, tap the submode name at the top of the screen (outlined in red in
customize the submode name.
4. Select the pump type (
Peristaltic
or
Venturi
).
5. Use the up and down arrows to set the
Aspiration
rate (0 to 60).
6. Use the up and down arrows to set the
Vacuum
. If you are using Venturi, the range is 0 to
600. If you are using Peristaltic, the range is 0 to 650.
7. Select the delivery mode (
Linear
or
Panel
).
8. Set the
Bottle Height
(Advanced Fluidics) or
Total Height
(Advanced Infusion).
9. Tap
Save
to save the program.
Содержание Veritas
Страница 1: ...Z370584 Rev D 09 2020 Johnson Johnson Surgical Vision Inc 2020 Software Release 8 000 OPERATOR MANUAL...
Страница 2: ...Z370584 Rev D This page was intentionally left blank...
Страница 4: ...Z370584 Rev D This page was intentionally left blank...
Страница 22: ...Z370584 Rev D 1 12 This page was intentionally left blank...
Страница 36: ...Z370584 Rev D 2 14 This page was intentionally left blank...
Страница 54: ...Z370584 Rev D 3 18 This page was intentionally left blank...
Страница 122: ...Z370584 Rev D 5 42 This page was intentionally left blank...
Страница 124: ...Create a New Program Z370584 Rev D 6 2 Program Settings Figure 6 1 Surgeons and Programs Screen 1 On Select Program tap...
Страница 146: ...WHITESTAR Technology Z370584 Rev D 6 24 Program Settings Figure 6 14 Duty Cycles for Variable WS...
Страница 158: ...Z370584 Rev D 6 36 This page was intentionally left blank...
Страница 184: ...Z370584 Rev D 7 26 This page was intentionally left blank...
Страница 198: ...Z370584 Rev D 9 4 This page was intentionally left blank...
Страница 270: ...Z370584 Rev D I 8 This page was intentionally left blank...