Warnings, Cautions, and Other Residual Risks
Z370584 Rev. D
Warnings
2-7
•
This equipment/system is intended for use by healthcare professionals only. This
equipment/system may cause radio interference or may disrupt the operation of nearby
equipment. It may be necessary to take mitigation measures, such as re-orienting or
relocating the VERITAS Vision System or shielding the location.
•
The VERITAS Vision System needs special precautions regarding electromagnetic
compatibility (EMC) and needs to be installed and put into service according to the EMC
information provided in this manual.
•
Portable and mobile RF communications equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 30 cm (12 inches) to any part
of the VERITAS Vision System, including cables specified by JJSV. Otherwise, degradation
of the performance of this equipment could result.
•
The use of accessories, transducers and cables other than those specified by JJSV, may
result in increased electromagnetic EMISSIONS or decreased electromagnetic IMMUNITY
of the VERITAS Vision System.
•
The VERITAS Vision System should not be used adjacent to or stacked with other
equipment and that if adjacent or stacked use is necessary, the VERITAS Vision System
should be observed to verify normal operation in the configuration in which it will be used.
•
Do not replace the foot pedal battery when the pedal is attached to a power source.
•
The VERITAS Vision System may be interfered with by other equipment, even if that other
equipment complies with regulatory emission requirements such as the International
Special Committee on Radio Interference (CISPR).
•
Inadvertent activation of functions that are intended for priming or tuning handpieces
while the handpiece is in the eye can create a hazardous situation that could result in
patient injury.
•
Danger of ignition of endogenous gases (e.g., cotton and gauze saturated with oxygen may
be ignited by sparks produced during normal use of HIGH FREQUENCY (HF) surgical
equipment).
•
Ensure the maximum capacity of the drain container is not exceeded as this could cause a
hazardous situation to the patient.
•
The ophthalmic irrigation solution source shall be at or above the patient's eye level.
•
Never intentionally modify handpieces or tips (e.g., do not bend, cut, or engrave them) as
they could break or malfunction.
•
Use caution when using clamps on a tubing set.
•
Use only recommended tubing set(s).
•
Do not touch an activated ultrasonic handpiece tip, as injuries could occur.
Содержание Veritas
Страница 1: ...Z370584 Rev D 09 2020 Johnson Johnson Surgical Vision Inc 2020 Software Release 8 000 OPERATOR MANUAL...
Страница 2: ...Z370584 Rev D This page was intentionally left blank...
Страница 4: ...Z370584 Rev D This page was intentionally left blank...
Страница 22: ...Z370584 Rev D 1 12 This page was intentionally left blank...
Страница 36: ...Z370584 Rev D 2 14 This page was intentionally left blank...
Страница 54: ...Z370584 Rev D 3 18 This page was intentionally left blank...
Страница 122: ...Z370584 Rev D 5 42 This page was intentionally left blank...
Страница 124: ...Create a New Program Z370584 Rev D 6 2 Program Settings Figure 6 1 Surgeons and Programs Screen 1 On Select Program tap...
Страница 146: ...WHITESTAR Technology Z370584 Rev D 6 24 Program Settings Figure 6 14 Duty Cycles for Variable WS...
Страница 158: ...Z370584 Rev D 6 36 This page was intentionally left blank...
Страница 184: ...Z370584 Rev D 7 26 This page was intentionally left blank...
Страница 198: ...Z370584 Rev D 9 4 This page was intentionally left blank...
Страница 270: ...Z370584 Rev D I 8 This page was intentionally left blank...