Surgery Start Up
Z370584 Rev. D
Handpiece Setup
4-7
Phacoemulsification Handpiece Setup – WHITESTAR and
ELLIPS FX
1.
Use caution to prevent burns when handling the handpiece directly from sterilization.
2. Remove the tubing pack and accessories from the packaging and place them in the
sterile area.
3. Assemble the phaco handpiece as shown below. You need the handpiece, phaco tip,
appropriate tip wrench, one of the irrigation sleeves, and the test chamber.
Figure 4-6 — Phaco Handpiece Assembly
4. Attach the connector end of the handpiece to the phaco receptacle on the front of the
and
).
5. Ensure there is no moisture on the connectors prior to attaching. Moisture prevents the
handpiece from operating properly.
WARNING!
Sterility assurance is the responsibility of the user. You must sterilize all non-sterile
accessories prior to use.
WARNING!
Prior to using any invasive portions of the handpiece assembly, examine under the
microscope for any obvious damage, oxidation, or the presence of foreign material.
Use a backup handpiece for surgery if there are any questionable characteristics of
the handpiece. Use of contaminated or damaged system accessories can cause
patient injury.
CAUTION!
Never attempt to straighten a bent phaco ultrasonic tip. This might produce a
broken tip when you apply ultrasound.
1. Test Chamber
2. Irrigation Sleeve
3. Phaco Tip
4. Handpiece
Содержание Veritas
Страница 1: ...Z370584 Rev D 09 2020 Johnson Johnson Surgical Vision Inc 2020 Software Release 8 000 OPERATOR MANUAL...
Страница 2: ...Z370584 Rev D This page was intentionally left blank...
Страница 4: ...Z370584 Rev D This page was intentionally left blank...
Страница 22: ...Z370584 Rev D 1 12 This page was intentionally left blank...
Страница 36: ...Z370584 Rev D 2 14 This page was intentionally left blank...
Страница 54: ...Z370584 Rev D 3 18 This page was intentionally left blank...
Страница 122: ...Z370584 Rev D 5 42 This page was intentionally left blank...
Страница 124: ...Create a New Program Z370584 Rev D 6 2 Program Settings Figure 6 1 Surgeons and Programs Screen 1 On Select Program tap...
Страница 146: ...WHITESTAR Technology Z370584 Rev D 6 24 Program Settings Figure 6 14 Duty Cycles for Variable WS...
Страница 158: ...Z370584 Rev D 6 36 This page was intentionally left blank...
Страница 184: ...Z370584 Rev D 7 26 This page was intentionally left blank...
Страница 198: ...Z370584 Rev D 9 4 This page was intentionally left blank...
Страница 270: ...Z370584 Rev D I 8 This page was intentionally left blank...