![Instramed CardioMax Скачать руководство пользователя страница 83](http://html.mh-extra.com/html/instramed/cardiomax/cardiomax_user-manual_2070096083.webp)
User manual |
Defibrillator mode
83
Result screens for functional tests
Remember: If CardioMax fails the functional test, contact technical support
IMMEDIATELY.
NOTE:
printouts of test results will only be available in the CardioMax units
equipped with a thermal printer.
NOTE:
the CardioMax indicates failure in the functional test when there is a failure
in one of the 4 steps of the functional test or when the power delivered has an
energy exceeding that allowed by health standards.
Put pads on the base
Select 100J
Press CHARGE
Press SHOCK
Passed test!
FUNCTIONAL TEST
OK
Functional test
Passed test!
exit
back
Put pads on the base
Selecit 100J
Press CHARGE
Press SHOCK
Failed test!
FUNCTIONAL TEST
OK
Functional test
Failed test!
Contact technical support!
exit
back
Equipment passed functional test
Equipment failed functional test