![Instramed CardioMax Скачать руководство пользователя страница 143](http://html.mh-extra.com/html/instramed/cardiomax/cardiomax_user-manual_2070096143.webp)
User manual |
RTC - Real Time Check
143
Manual test
To open the dialog box for the manual test, access CONFIGURATION > GENERAL
CONFIGURATIONS > RTC > PERFORM TEST MANUALLY.
Put the Paddles on the support
□
□
Select 30J
cancel
Next
Manual test
Follow the instructions on the screen and click on "next" to start the test. Or click on
"cancel" to close the dialog box and return to the RTC menu.
Once started, the manual test goes through the following steps:
• Test of the connection with the ECG board.
• Test of the defibrillator module.
• Check of the battery level.
• Check of the connection with the power outlet.
□
□
□
WARNING
RTC MANUAL TEST
98%
Test completed
Sending data to the Central. Wait...
Communication with the ECG
Defibrillation - 30J
Battery level
Connected to the main power
WAIT