User manual |
Defibrillator mode
80
Charge Auto-Sequencing mode (Auto Seq)
There are three sequential energy levels, being that, from the third shock on all
subsequent shocks will use the same discharge value of the latter. The preset
sequence will be interrupted in the following conditions:
• Equipment shutdown.
• Auto Seq mode exit.
• Reconfiguration of the Auto-Sequencing mode energy levels (available after the
delivery of the third shock).
If child pads are used, the equipment will automatically limit the charge value to 50 J.
Auto Sequencing
50 J
100 J
200 J
50:45
Adult
0
The energy to
be charged is
indicated in red.
Auto Sequencing
50 J
100 J
200 J
50:45
Adult
1
After shock
delivery, the
energy is
updated with
the next energy
level.
Monitor
Energy
OFF
Seq.
Auto
Pacemaker
In the Auto Seq mode, the shock energy
levels will follow the order previously set by
the user in: "Defibrillation > Auto Sequencing
Charge” (See “Defibrillation setup” below).