After completing the test, the equipment will send the collected data to the RTC
Control Center. At this time, it is not possible to turn off the device. If the connection
with the RTC Control Center cannot be established, the device will cancel sending
the data after 10 seconds. After this procedure, the device will save the collected
data and turn off automatically.
□
□
□
WAIT
98%
WARNING
RTC AUTO TEST
Test completed
Sending data to the Central. Wait...
Communication with the ECG
Defibrillation - 30J
Battery level
Connected to the main power
Note: The test of the defibrillator module will not be carried out if:
•
The equipment is connected to the computer by USB cable.
•
The paddles are not on the support or are disconnected from the device
(External Adult Paddles).
•
The battery has a charge of less than 10% and the equipment is not
connected to a power outlet.