User manual |
Alarms and limits
58
Tachycardia (↓)
HIGH PRIORITY
RED
When the bpm is below the value set on the alarm
configuration.
Asystole
HIGH PRIORITY
RED
CardioMax cannot detect valid beats for more than
4 seconds.
Shock Indicated
HIGH PRIORITY
RED
Ventricular fibrillation / ventricular tachycardia(VF /
VT) was identified.
RESPIRATION
ALARM
PRIORITY
COLOR
POSSIBLE CAUSE
Apnea Alarm
HIGH PRIORITY
RED
When detected the suspension of breathing
(apnea) on specified times 5, 10, 15, 20, 25, 30,
35 or 40 seconds (set at Menu > RESP > Apnea
Alarm).
RPM High(↑)
HIGH PRIORITY
RED
When the RPM is above the value defined on the
alarm configuration.
RPM Low(↓)
HIGH PRIORITY
RED
When the BPM is below the value defined on the
alarm configuration.
SpO
2
ALARM
PRIORITY
COLOR
POSSIBLE CAUSE
No finger on the
sensor
MEDIUM PRIORITY
Yellow
Connected sensor on the equipment, but without
detection of patient’s finger.
Searching signal
MEDIUM PRIORITY
Yellow
The monitor is searching for valid SpO
2
signal.
SpO
2
–
Disconnected
Sensor
MEDIUM PRIORITY
Yellow
Sensor or SpO
2
extension disconnected or
incorrectly placed.
SpO
2
- Artifact
MEDIUM PRIORITY
Yellow
Muscle tremor detected.
Weak signal
MEDIUM PRIORITY
Yellow
CardioMax cannot identify the signal. Weak signal,
possibly patient with low perfusion.
Loss of pulse
HIGH PRIORITY
RED
Patient without heartbeats for more than 4
seconds
Saturation
High (↑)
HIGH PRIORITY
RED
Saturation is above the value defined on the alarm
configuration.
Saturation
Low (↓)
HIGH PRIORITY
RED
Saturation is below the value defined on the alarm
configuration.