![Instramed CardioMax Скачать руководство пользователя страница 106](http://html.mh-extra.com/html/instramed/cardiomax/cardiomax_user-manual_2070096106.webp)
User manual |
Monitor mode - ECG
106
3 - Sensitivity
Select the ECG amplification gain. You can select from 5, 10, 15, 20, 30 or
40 mm/mV.
4 - Frequency bandwidth selection
Filter selection for power interference. “Diag” or “Monitor”.
5 - Mains supply filter
Enables to turn filter on or off to reduce mains supply interference with the ECG
signal.
6 - Detect pacemaker
Enables to turn the pacemaker detection mode on or off. When on, the equipment
indicates on the screen the pacemaker’s pulse moment. Detect pacemaker mode
must only be used on patients with pacemaker.
7 - ST segment
Enables/disables the presentation of the ST segment. The visual indication will
occur in the column at the end of the parameter curves, before the numerical
values.
8 - Alarm
The “BELL” icon indicates audio “OFF”, “ON” or “PAUSED”. Alarm limits
configuration “MINIMUM” and “MAXIMUM”.
9 - Back/Exit
“Back” to Configuration menu and “Exit” to monitoring screen.