Defibrillator mode
73
About shock delivery
Aligning the pressure of the pads and the conductive material applied to the
electrodes, different patient impedances are obtained.
The table below indicates the conditions in which the CardioMax offers or inhibits
the delivery of energy.
Patient's
impedance
Shock
Message on screen after
“Charge” key pressed
Values indicated on
bargraph
Short circuit
Shock inhibited
Bad contact
All LEDs blinking
< 25 Ohm
Shock inhibited
Bad contact
All LEDs blinking
>25 Ohm e
< 300 Ohm
Shock delivered
and the waveform is
adjusted according
to the patient's
impedance
No message
LEDs lit up indicating contact
level
> 300 Ohm
Shock inhibited
Bad contact
Only the red LED is blinking
Short - open
Shock inhibited
Bad contact
Only the red LED is blinking
When all LEDs blink simultaneously the pads have a short circuit, and
shock delivery will not be permitted.
When only the RED LED is blinking, shock delivery will not be allowed.