Alarms and limits
59
NIBP
ALARM
PRIORITY
COLOR
POSSIBLE CAUSE
Excessive
Pressure
MEDIUM PRIORITY
YELLOW
The cuff´s maximum pressure has been
exceeded.
Cuff problems
MEDIUM PRIORTY
YELLOW
The cuff is misplaced or there is leakage on
measure circuit.
Weak signal
MEDIUM PRIORITY
YELLOW
Pulse too weak for NIBP measurements. Check
the cuff´s position and tightening.
Excessive
movement
MEDIUM PRIORITY
YELLOW
There is noise due to the patient’s movement.
Long
measurement
MEDIUM PRIORITY
YELLOW
The measurement of pressure is too long with
possibility of inaccurateness.
Pressure
(Systolic,
Diastolic or
Medium) High
(↑)
HIGH PRIORITY
RED
The pressure is above the value defined on the
alarm configuration.
Pressure
(Systolic,
Diastolic or
Medium) Low (↓)
HIGH PRIORITY
RED
The pressure is below the value defined on the
alarm configuration.
EtCO
2
ALARM
PRIORITY
COLOR
POSSIBLE CAUSE
No Filterline
MEDIUM PRIORITY
YELLOW
The capnography sampling line is not connected.
Occlusion
MEDIUM PRIORITY
YELLOW
There is no airflow in the EtCO
2
sensor. Change
the sampling line (Filterline).
Starting sensor
MEDIUM PRIORITY
YELLOW
The EtCO
2
modulo is heating up its internal
sensors (this occurs during the beginning of
capnography and lasts about 15 seconds).
Auto-Zero
MEDIUM PRIORITY
YELLOW
Necessary procedure for proper functioning of the
equipment.
CO
2
EXP High(↑)
HIGH PRIORITY
RED
CO
2
EXP is above the value defined on the alarm
configuration.
CO
2
EXP Low(↓)
HIGH PRIORITY
RED
CO
2
EXP is below the value defined on the alarm
configuration.
Communication
failure
HIGH PRIORITY
RED
Communication failure of EtCO
2
module.