User manual |
Defibrillator mode
75
Using multifunction pads
The use of the multifunction pads (disposable) requires that the operator uses the
adapter provided (extension cable) in order to connect them to the standard socket
of the external pads, as described below:
1
- Connect the AED/pacemaker extension cable to the equipment.
2
- Connect the adhesive multifunction pads to the extension cable.
3
- Remove the multifunction pads’ protective film and apply them to the patient,
using the same positions recommended for the external pads (Sternum and Apex).
ATTENTION: The pads are disposable and for single-use, cannot be reused
under any circumstances.