User manual |
Alarms and limits
65
Auto set
The AUTO SET function configures alarm limits taking into account the physiological
parameter values that are instantly measured on the patient by calculating a
deviation of the minimum and maximum limits. See the table below:
Parameter
Minimum
Maximum
ECG
X 0.8
X 1.6
SpO
2
Default
Default
NIBP - Systole
X 0.7 + 10
X 0.9 + 40
NIBP - Diastole
X 0.7 + 6
X 0.9 + 34
NIBP - Mean
X 0.7 + 6
X 0.9 + 35
Respiration
Default
Default
EtCO
2
Default
Default
For instance, if a patient registers a cardiac frequency of 60 BPM, the values for the
AUTO-ADJUSTMENT function will be: Minimum = 48 and Maximum = 96.
Silence
It allows the silence time of the physiological and technical alarms to be adjusted
with times between 30 and 180 seconds. When the “Alarm Silenced” indication is
ON, the silence time adjustment is blocked until the indication is back to normal.
Auto set
30
180
[180 ]
ALARM SETUP
85
100
SPO2
20
60
CO2
6
45
90
sis
200
50
dia
150
50
med
170
40
160