User manual |
Monitor mode - ECG
105
ECG Setup
WAVEFORM
ECG Response
Slow
Fast
[Normal
]
ECG Cable
3-wire
10-wire
[5-wire
]
Detect Pacemaker
Off
On
[Off
]
Segmento ST
Off
On
[Off
]
Sensitivity
5
40
[20
]
Frequency bandwith
Diag
Monitor
[Monitor ]
Mains supply filter
Off
On
[On
]
40
160
ECG
ECG SETUP
back
exit
-4.4
10.0
ST
1
3
4
7
6
2
5
8
9
1 - ECG response
Select the ECG numeric update response. You can select from “SLOW, “NORMAL”
and “FAST” responses.
NORMAL:
used for most patients, this mode uses 16 beats to define the average.
FAST:
used when the operator needs faster responses. It is affected by the patient's
movements, using 8 beats to define the average.
SLOW:
less affected by the patient's movements. However, you must pay attention
to the slow response of the heart rate variation. This mode uses 32 beats to define
the average.
2 - ECG cable
It allows selecting the ECG monitoring by cable with 3 lead, 5-lead or 10-lead.
Using the 3-wire ECG cable in the SDP (Sudden Death Prevention) makes
it functional only in the DII lead.