Troubleshooting
21
182
Symptom
Probable cause
Probable solution
The CardioMax does not turn on.
There is no electricity.
- Check connections: the CardioMax/Power
Cable/Plug.
Does not select energy > 50 J.
Adult pads identification.
- Check if is adult pads are connected to the
equipment and if adult electrodes are properly
connected.
Does not deliver shock.
Impedance measuring.
- Check the graph bar for the patient’s
impedance indication.
Does not capture ECG via
electrodes.
Lead selection.
- Select a lead other than the “pads” lead.
No tracing.
Unstable since last
shutdown.
- Restore initial configuration.
Low battery autonomy.
Defective battery.
- Turn off battery.
No QRS audio indication.
BEEP volume.
- Turn on BEEP’s volume in the configuration
menu.
No alarm sound indications.
Operation mode.
- Alarm indications are only active in the monitor
mode.
No SpO
2
curve.
SpO
2
curve turned off.
- Turn on SpO
2
curve in Waveform setup.
Pacemaker does not start.
Multifunction/Adhesive
pads.
- Check if multifunction pads are connected.
- Check if there is a “bad contact” message.
- Check gel in multifunction pads.
Does not print .
No paper in the printer.
- Check if there is paper in the printer.
- Check if paper is positioned properly.
Printer makes noises and does
not print.
Too much paper on roll.
- Change roll’s size.