Monitor mode - Respiration
129
Respiration setup
Using the e-Jog select the respiration function in the setup menu to access the
respiration setup sub-menu.
WAVEFORM
Sensitivity
[6 ]
1
6
RESP SETUP
Resp On/Off
Respiratory Frequency
Apnea Alarm
[Off
]
[ECG
]
[Off
]
Off
ECG
Off
On
CO2
40
back
exit
1
2
3
4
5
1 - Respiration monitoring On/Off
When turned off, all visual and sound alarms will be inhibited and there will be no
numeric indication for respiration.
2 - Respiratory frequency
Determines if the frequency shown by the equipment will be obtained by thoracic
impedance (ECG cable) or by capnography (CO
2
).
3 - Apnea alarm
The alarm emits a sound when a suspension of breathing is detected (apnea) in the
specified times of 5, 10, 15, 20, 25, 30, 35 or 40 seconds.
4 - Respiration sensitivity
You can select among the following values:1, 2, 3, 4, 5 and 6.