User manual |
Monitor mode - SpO
2
117
Prolonged use or the condition of the patient may require replacement of
the sensor. Replace the sensor every 4 hours and often check the integrity
and circulatory condition of the skin as well as the correct alignment of the
sensor.
If the finger cannot be positioned correctly, choose a smaller finger or use
another sensor.
Factors which affect the SpO
2
measurement’s precision
• Incorrect use of the sensor.
• Anemia.
• Use of vasoactive drugs.
• Patient in shock or cardiac arrest.
• Significant number of dysfunctional hemoglobins.
• Intravascular contrasts such as indocyanine green and methylene blue.
• Arterial occlusion next to the sensor.
Sensor selection
Choose an appropriate sensor in the chart below. Check the instructions which
accompany the sensor for application instructions.
Patient
Place
Description
Adult/ Pediatric
Finger (hand).
Finger or toe (hand or foot).
12556: Sensor, adult. Ref: 3044.
12475: Sensor, “Y” universal. Ref: 3043.
Child
Finger or foot.
12475: Sensor, “Y” universal. Ref: 3043.