![Instramed CardioMax Скачать руководство пользователя страница 145](http://html.mh-extra.com/html/instramed/cardiomax/cardiomax_user-manual_2070096145.webp)
User manual |
RTC - Real Time Check
145
RTC Operation on the CardioMax 8 without transmitter
The equipment that do not have RTC module installed, will not have communication
with PC. The functionalities of the RTC will be the same as listed above, without
performing data transmission. After the manual or automatic test, the equipment
will save the collected data and will put it available for visualization on equipment
through the option “Last test status”.
Daily tests quantity
3
06:00
12:00
20:00
First Schedule
Second Schedule
Third Schedule
Perform test manually
Last test status
exit
RTC (no transmitter)
1. Number of automatic daily tests. It can be
configured with values of 1 to 3 tests daily.
2. Time for performing the first daily test.
Configured by the user.
3. Time for performing the second daily test.
Configured by the user. This item will be
disabled when the number of daily tests is 1.
4. Time for performing the third daily test.
Configured by the user. This item will be
disabled when the number of daily tests is
less than 3.
5. Carry out a manual test, without the need for
prior configuration.
6. Displays the data from the last test
performed.