User manual |
Specifications and safety
193
Precision of applied energy:
Energy
selected
Impedance
Accuracy
25
50
75
100
125
150
175
1
0.9
1.0
1.1
1.1
1.0
0.9
0.8
±
1J
2
1.8
1.9
2.0
2.0
2.0
2.0
1.9
±
1J
3
2.8
3.0
3.0
3.0
3.1
3.2
3.2
±
2J
4
3.6
3.9
3.9
4.0
4.0
3.9
3.9
±2J
5
4.8
5.1
5.1
5.0
5.0
5.0
4.9
±3J
6
5.5
5.8
5.9
6.0
6.0
6.0
6.0
±3J
7
6.5
6.9
7.2
7.2
7.1
7.0
7.0
±3J
8
7.2
7.9
8.1
8.2
8.3
8.1
7.7
±3J
9
7.8
8.6
8.9
9.0
9.0
9.0
8.8
±3J
10
8.8
9.8
10.2
10.4
10.3
10.2
9.8
±3J
20
19.0
20.5
21.0
21.0
20.5
19.5
19.0
±15%
30
27.5
30.0
31.0
31.5
31.0
29.5
27.5
±15%
50
49.0
52.0
53.0
52.5
51.5
48.0
45.5
±15%
80
77.5
81.5
82.5
83.0
80.5
76.5
74.5
±15%
100
96.0
101.0
102.5
103.5
101.0
96.5
92.0
±15%
150
143.0
151.5
155.0
153.0
148.0
141.0
137.0
±15%
200
191.5
201.5
205.5
206.0
203.5
192.0
177.0
±15%
250
240.0
250.5
256.5
256.0
254.0
241.5
224.0
±15%
300
284.0
302.0
305.5
306.0
305.0
290.0
270.0
±15%
360
344.0
363.0
370.5
370.0
363.0
345.0
322.0
±15%
Patient's impedance response table:
Patient Impedance
Shock
Short-circuit
Shock inhibited
< 25 Ohms
Shock inhibited
> 25 Ohms AND < 300 Ohms
Shock delivered with a waveform adjusted to the
patient's impedance
> 300 Ohms
Shock inhibited
Open circuit
Shock inhibited