MediaClave overview
INTEGRA Biosciences
Page 52
7.3
Country versions
7.3.1
Electrical ratings
7.4
Software default values
7.4.1
Overview
This section informs about the different default program settings which comes with a new MediaClave.
Furthermore, the default general values, which can be changed in the
PARAMETER
Menu, are listed.
7.4.2
MEDIACLAVE programs 1–13 (1–9 L of medium)
WARNING
:
For all countries worldwide, a protective earth (PE) wire is required
NOTE:
Mains voltage tolerance: +/– 10%.
Region
Mains voltage
(VAC)
Phases
min.
current (A)
Wattage
(kW)
Frequency
(Hz)
Article
Number
Europe
3 x 400
3L,N,PE
6A per Ph.
3.5
50–60
135030
Australia,
Europe, Africa
1 x 230
L,N,PE
16A
3.5
50–60
135020
USA
1 x 240 (2x120)
2L,PE,180°
16A
3.5
50–60
135025
Japan
1 x 200
L,N,PE
20A
3.5
50–60
135035
Table 7.2:
Fixed voltage
PROGRAMS 1–13
Range
Default
Sterilisation temperature:
70 - 122°C
121°C
Sterilisation time:
1 - 99 min.
20 minutes
Stirrer speed:
1,2
2 (150 rpm)
Dispensing temperature:
30 - 80°C
50°C
Stirrer speed:
1,2
1 (75 rpm)
Table 7.3:
MEDIACLAVE programs 1 - 13
Содержание MediaClave
Страница 1: ...User Manual 3 4 Bedienungsanleitung 3 4 Mode d emploi 3 4...
Страница 6: ...Table of contents INTEGRA Biosciences PAGE IV...
Страница 12: ...MediaClave overview INTEGRA Biosciences Page 6...
Страница 18: ...MediaClave overview INTEGRA Biosciences Page 12...
Страница 30: ...MediaClave overview INTEGRA Biosciences Page 24...
Страница 48: ...MediaClave overview INTEGRA Biosciences Page 42...
Страница 56: ...MediaClave overview INTEGRA Biosciences Page 50...
Страница 68: ...MediaClave overview INTEGRA Biosciences Page 62...
Страница 70: ...MediaClave overview INTEGRA Biosciences Page 64...
Страница 74: ...MediaClave overview INTEGRA Biosciences Page 68...