User Manual 3.4
INTEGRA Biosciences
Page 41
ENGLISH
)
•
VESSEL IS BEING HEATED
•
VESSEL IS BEING CLEANED
•
VESSEL IS BEING EMPTIED
(The actual temperature also appears)
3.
Clean and empty the vessel as outlined in section 5.3.3.2, steps 2–5
NOTE:
If there is insufficient water in the vessel, the following alarmed message appears –
TOO LITTLE HEATING WATER IN VESSEL
. Press
END
to confirm this error.
NOTE:
The actual and target temperatures are displayed in the upper right of the screen. The pre-
programmed target temperature (see Section 3.4.4.2) must be achieved in order for the cleaning cy-
cle to progress.
NOTE:
The countdown timer is displayed in the upper right of the screen. The temperature is held at
the target value for 100 seconds. The timer counts back from 100 to 0 seconds.