User Manual 3.4
INTEGRA Biosciences
Page 15
ENGLISH
3.2.2
Mode AUTOCLAVE
3.2.2.1
Installing the flexible temperature sensor
1.
Unplug the rigid temperature sensor form the rear panel if installed
2.
Using an Allen tool, unscrew the rigid temperature sensor from the vessel lid
3.
Insert the flexible temperature sensor, and tighten with the Allen tool
4.
Plug in to the rear panel
3.2.2.2
Installing the steel base plate
1.
Remove the cuvette
2.
Place the steel base plate into the vessel, flat side facing up
3.2.2.3
Filling the vessel
In mode AUTOCLAVE, fill the vessel with distilled or deionised water to a height of 2 cm above the insert-
ed base plate (this corresponds to a height of 3.5 cm without the inserted base plate).
NOTE:
The vessel lid must be closed at all times during operation of the AUTOCLAVE programs!
ATTENTION:
Ensure that there is a seal attached to the temperature sensor.
NOTE:
If insufficient water is added, an alarmed error message appears on the screen. Follow all on-
screen instructions.
Содержание MediaClave
Страница 1: ...User Manual 3 4 Bedienungsanleitung 3 4 Mode d emploi 3 4...
Страница 6: ...Table of contents INTEGRA Biosciences PAGE IV...
Страница 12: ...MediaClave overview INTEGRA Biosciences Page 6...
Страница 18: ...MediaClave overview INTEGRA Biosciences Page 12...
Страница 30: ...MediaClave overview INTEGRA Biosciences Page 24...
Страница 48: ...MediaClave overview INTEGRA Biosciences Page 42...
Страница 56: ...MediaClave overview INTEGRA Biosciences Page 50...
Страница 68: ...MediaClave overview INTEGRA Biosciences Page 62...
Страница 70: ...MediaClave overview INTEGRA Biosciences Page 64...
Страница 74: ...MediaClave overview INTEGRA Biosciences Page 68...