![IBS MediaClave Скачать руководство пользователя страница 51](http://html1.mh-extra.com/html/ibs/mediaclave/mediaclave_user-manual_3298851051.webp)
User Manual 3.4
INTEGRA Biosciences
Page 45
ENGLISH
6.3
Temperature calibration
6.3.1
Overview
Company directives require all companies working under a quality assurance management to periodically
calibrate their equipment. The calibration must guarantee that the instrument fulfils the defined specifica-
tions.
6.3.2
When to perform calibration
INTEGRA Biosciences recommends the following:
•
Calibration must be performed at least once a year, preferably in combination with a service
•
Calibration must be performed after each instrument adjustment or repair
6.3.3
Temperature calibration routine
For temperature calibration, an external temperature measuring system is required. Instrument and sen-
sor must be calibrated together. To ensure an high accuracy, a good temperature instrument and a high-
class sensor is needed. INTEGRA Biosciences provides an appropriate system including a temperature-
measuring instrument and a double Pt100 temperature sensor 1/10 DIN with a calibration protocol for the
entire system. (Order number: 135800)
6.3.3.1
Using the INTEGRA temperature calibration set (Order Nr. 135800)
•
Connect the double Pt100 sensor to the temperature sensor inlet. There are two connector cables.
One is connected to the MediaClave rear panel, the other one to the external instrument.
•
Compare the true temperature reading with the temperature reading on the display. If the displayed
reading differs from the true reading more than defined by internal company directives, the tempera-
ture measurement must be adjusted. Contact a Service Technician (see chapter 8).
6.3.3.2
Using another temperature measurement system for calibration
•
An external temperature measuring system, consisting of a measuring instrument and a Pt100 tem-
perature sensor, is required.
•
Connect this system to the dispense port
•
Compare the reference temperature reading with the temperature reading on the display. If the dis-
played reading differs from the reading, contact a Service Technician (see chapter 8 for addresses).
ATTENTION:
Reference temperature instrument and sensor must be calibrated together as a sys-
tem.
ATTENTION:
To make sure that there is a homogeneous temperature inside the MediaClave, the
stirrer must be switched on.
Содержание MediaClave
Страница 1: ...User Manual 3 4 Bedienungsanleitung 3 4 Mode d emploi 3 4...
Страница 6: ...Table of contents INTEGRA Biosciences PAGE IV...
Страница 12: ...MediaClave overview INTEGRA Biosciences Page 6...
Страница 18: ...MediaClave overview INTEGRA Biosciences Page 12...
Страница 30: ...MediaClave overview INTEGRA Biosciences Page 24...
Страница 48: ...MediaClave overview INTEGRA Biosciences Page 42...
Страница 56: ...MediaClave overview INTEGRA Biosciences Page 50...
Страница 68: ...MediaClave overview INTEGRA Biosciences Page 62...
Страница 70: ...MediaClave overview INTEGRA Biosciences Page 64...
Страница 74: ...MediaClave overview INTEGRA Biosciences Page 68...