User Manual 3.4
INTEGRA Biosciences
Page 27
ENGLISH
4.3.1.7
Dispensing the product
During the dispensing phase, the product will be dispensed through the sterile dispensing tube by an ex-
ternal pump.
1.
Unscrew the media outlet cap from the dispensing ports (see figure 2.1)
2.
Connect the sterile dispensing tube. When using a rigid dispensing tube,
flame briefly with a bunsen burner before inserting
3.
Secure the tube with the securing nuts
4.
Place the tube in the external pump
5.
When the cuvette is empty, press END
6.
Remove the dispensing tube from the dispensing nipple
7.
Clean the media outlet
8.
Close the outlet by covering with the cleaned cap
9.
Open the sterilisation compartment
10. Clean the rubber seal and insert again
11. Remove the cuvette (see section 3.2)
12. Clean the cuvette (see section 5.2)
13. Disconnect and clean the outlet tube immediately (see section 5.2).
•
PROGRAM 14 (Without pressure test) results in pressure free steam disinfection of growth media.
As the name suggests, a pressure test is not performed at the beginning of the cycle (true sterilisation
is not performed).
•
PROGRAM 15 (Chocolate Agar) is used for the production of chocolate agar. It runs in two cycles
(sterilisation and boiling).
Figure 4.1:
The MEDIACLAVE PROGRAM [#] Screen
Figure 4.1 illustrates the progress of the sterilisation cycle as displayed on the MEDIACLAVE PROGRAM
[#] screen.
4.3.1.8
Selecting programs
1.
Press
MEDIACLAVE
on the start display (see section 3.4.1). The MEDIACLAVE PROGRAM [#]
screen appears (see figure 4.1)
2.
Press
PROGRAM
. The PROGRAM menu appears (see figure 3.4)
3.
Press
PROG –
or
PROG +
to select the desired program
4.
Press
BACK
to transfer the selected program to the MEDIACLAVE PROGRAM [#] screen
5.
Press
START
to start the selected program (see section 4.3.1.10)
ATTENTION:
MEDIACLAVE programs 14 and 15 differ from the other 13 programs:
ating
Securing nut
Содержание MediaClave
Страница 1: ...User Manual 3 4 Bedienungsanleitung 3 4 Mode d emploi 3 4...
Страница 6: ...Table of contents INTEGRA Biosciences PAGE IV...
Страница 12: ...MediaClave overview INTEGRA Biosciences Page 6...
Страница 18: ...MediaClave overview INTEGRA Biosciences Page 12...
Страница 30: ...MediaClave overview INTEGRA Biosciences Page 24...
Страница 48: ...MediaClave overview INTEGRA Biosciences Page 42...
Страница 56: ...MediaClave overview INTEGRA Biosciences Page 50...
Страница 68: ...MediaClave overview INTEGRA Biosciences Page 62...
Страница 70: ...MediaClave overview INTEGRA Biosciences Page 64...
Страница 74: ...MediaClave overview INTEGRA Biosciences Page 68...