User Manual 3.4
INTEGRA Biosciences
Page 43
ENGLISH
6
Maintenance
6.1
Introduction
The Maintenance Section consists of four parts:
•
Paper handling
•
Temperature calibration
•
Troubleshooting
•
Frequently asked questions FAQ‘s
6.2
Paper handling
6.2.1
Replacing the Printer paper roll
1.
Ensure that the MediaClave is switched on
2.
Remove the printer protection plexiglass and the magnetic front cover
3.
Press the form feed button to expel the last piece of paper
4.
Remove the empty bobbin and replace with a new paper roll
5.
Insert the end of the paper into the paper slit
6.
The paper is automatically fed into the printer
7.
Press the form feed button
8.
Replace the printer protection plexiglass and the magnetic front cover
9.
Perform a print self-test (see section 3.4.4.2 - PRINTER)
ATTENTION:
Refer to Figure 6.1 and Figure 6.2 when replacing the printer paper roll.
NOTE:
The appearance of a red strip at the edge of the paper indicates the need of a paper-replacement.
Содержание MediaClave
Страница 1: ...User Manual 3 4 Bedienungsanleitung 3 4 Mode d emploi 3 4...
Страница 6: ...Table of contents INTEGRA Biosciences PAGE IV...
Страница 12: ...MediaClave overview INTEGRA Biosciences Page 6...
Страница 18: ...MediaClave overview INTEGRA Biosciences Page 12...
Страница 30: ...MediaClave overview INTEGRA Biosciences Page 24...
Страница 48: ...MediaClave overview INTEGRA Biosciences Page 42...
Страница 56: ...MediaClave overview INTEGRA Biosciences Page 50...
Страница 68: ...MediaClave overview INTEGRA Biosciences Page 62...
Страница 70: ...MediaClave overview INTEGRA Biosciences Page 64...
Страница 74: ...MediaClave overview INTEGRA Biosciences Page 68...