User Manual 3.4
INTEGRA Biosciences
Page 29
ENGLISH
6.
Press
CONTINUE
to select the stirrer speed for boiling
7.
Press
–
or
+
to adjust the stirrer speed
8.
Press
CONTINUE
to select the cooling temperature
9.
Press – or + to adjust the cooling temperature
10. Press
CONTINUE
to select the stirrer speed for cooling
11. Press – or + to adjust the stirrer speed
12. Press
→
13. Press
SAVE
to save all modifications and return to the
PROGRAM
menu
OR
Press
CANCEL
to reject all modifications and return to the
PROGRAM
menu
14. Press
START
to start the selected program (see Section 4.3.1.10)
4.3.1.10
Starting and running programs
1.
Open the safety cover and vessel lid (see section 3.2)
2.
Press
MEDIACLAVE
on the start display. The MEDIACLAVE PROGRAM [#] screen appears
3.
Press
START
4.
Fill the cuvette with the medium to be prepared and enter the stirrer bar
5.
Install the cuvette in the vessel (see section 3.2)
6.
Install the rigid temperature sensor (see section 3.2)
7.
When dispensing with the dispensing tube (not with the rigid tube) connect this to the inner side of
the media outlet
8.
Close the vessel lid (see section 3.2)
9.
Press OK to continue the cycle
OR
Press
END
to terminate the cycle
10. Close the safety cover
.
•
PREPARE COOLING
appears on the GUI
•
All sensors are checked to ensure that they measure within the permitted temperature ranges. Sen-
sors are also checked for short-circuit, interruption and ice formation (< 2°C)
•
The pump runs for approximately five seconds to remove any trapped air
•
WATER CONTROL
appears. The water level is checked
•
PRESSURE TEST
appears for PROGRAMS 1–13 and 15. A pressure test is performed
•
The sterilisation cycle is started
NOTE:
The temperature can be adjusted between 30°C and 80°C.
ATTENTION:
Ensure there is sufficient heating water in the vessel. If there is insufficient heating water
in the vessel, the following alarmed message appears –
TOO LITTLE HEATING WATER IN VESSEL
.
Press END to confirm this error. The cycle is terminated.
ATTENTION:
Pressing
END
results in the alarmed message
CYCLE IS BEING CONCLUDED.
The
MAIN
menu appears.
NOTE:
The safety cover lock ensures that the safety cover is securely closed.
NOTE:
The heat exchanger is filled with water to prevent overheating. The actual and target temper-
atures are displayed in the upper right corner of the screen.
Содержание MediaClave
Страница 1: ...User Manual 3 4 Bedienungsanleitung 3 4 Mode d emploi 3 4...
Страница 6: ...Table of contents INTEGRA Biosciences PAGE IV...
Страница 12: ...MediaClave overview INTEGRA Biosciences Page 6...
Страница 18: ...MediaClave overview INTEGRA Biosciences Page 12...
Страница 30: ...MediaClave overview INTEGRA Biosciences Page 24...
Страница 48: ...MediaClave overview INTEGRA Biosciences Page 42...
Страница 56: ...MediaClave overview INTEGRA Biosciences Page 50...
Страница 68: ...MediaClave overview INTEGRA Biosciences Page 62...
Страница 70: ...MediaClave overview INTEGRA Biosciences Page 64...
Страница 74: ...MediaClave overview INTEGRA Biosciences Page 68...