User Manual 3.4
INTEGRA Biosciences
Page 21
ENGLISH
8.
Press
–
or
+
to perform the self-test
9.
Press
SAVE
to return to the PARAMETER ENTRY menu
COMMUNICATION
1.
From the PARAMETER ENTRY menu, press
CONTINUE
to scroll to and select COMMUNICATION.
2.
Press
EDIT
3.
The SET HOST ADDRESS menu displays two options:
•
DEVICE ADDRESS: [1–16]
•
HOST-PORT: 9600B, 1S, 8D, NP, 1S
4.
Press
–
or
+
to select the desired device address
5.
Press
SAVE
to return to the PARAMETER ENTRY menu
UNITS OF MEASUREMENT
1.
From the PARAMETER ENTRY menu, press
CONTINUE
to scroll to and select UNITS OF MEAS-
UREMENT
2.
Press
EDIT
3.
The SET UNITS OF MEASUREMENT menu displays the following:
•
UNITS OF MEASUREMENT: [DEGREES CELSIUS / DEGREES FAHRENHEIT]
4.
Press
–
or
+
to select the desired unit of measurement
5.
Press
SAVE
to return to the PARAMETER ENTRY menu
CONTRAST
1.
From the PARAMETER ENTRY menu, press
CONTINUE
to scroll to and select CONTRAST
2.
Press
EDIT
3.
The SET CONTRAST menu appears
4.
Press
–
or
+
to adjust the contrast to suit the working environment
5.
Press
SAVE
to return to the PARAMETER ENTRY menu
STERILISATION
1.
From the PARAMETER ENTRY menu, press
CONTINUE
to scroll to and select STERILISATION
2.
Press
EDIT
. The STERILISATION PARAMETER menu appears
3.
Press
CONTINUE
to choose MEDIACLAVE TOLERANCE or AUTOCLAVE TOLERANCE
4.
Press
–
or
+
to adjust the sterilisation tolerance
5.
Press
SAVE
to return to the PARAMETER ENTRY menu
NOTE:
A self-test is executed to check the functionality of the printer. When the self-test is initiated,
a test print is performed to check the functionality of the printer (see section 7.7).
NOTE:
A MediaClave device address can be selected from positions 1 to 16.
NOTE:
The contrast of the LCD can be adjusted between 0% and 100% to allow viewing in different light-
ing conditions.
NOTE:
The sterilisation tolerance can be adjusted between ±0.5°C and ±3.0°C. (default: ±1.5°C.)
Содержание MediaClave
Страница 1: ...User Manual 3 4 Bedienungsanleitung 3 4 Mode d emploi 3 4...
Страница 6: ...Table of contents INTEGRA Biosciences PAGE IV...
Страница 12: ...MediaClave overview INTEGRA Biosciences Page 6...
Страница 18: ...MediaClave overview INTEGRA Biosciences Page 12...
Страница 30: ...MediaClave overview INTEGRA Biosciences Page 24...
Страница 48: ...MediaClave overview INTEGRA Biosciences Page 42...
Страница 56: ...MediaClave overview INTEGRA Biosciences Page 50...
Страница 68: ...MediaClave overview INTEGRA Biosciences Page 62...
Страница 70: ...MediaClave overview INTEGRA Biosciences Page 64...
Страница 74: ...MediaClave overview INTEGRA Biosciences Page 68...